Warning: The indicated sterilization guidelines have been determined from sterilization
efficacy studies perfprmed on this or a related product and may be limited by
the process’s compatibility with the product’s specific materials of construction.
Sterilization should be performed following the parameters indicated, usingapproved
hospital/contractor procedures. The hospital/contractor is responsible for validating
all protocols and verifying the results of each sterilization process. All components
should be inspected prior to use.
CAUTION: Do not disassemble the PEEP Valve. Damage on these parts may occur
when disassembling them.
CLEANING GUIDELINES for the product: CLEAN-1 (Wash): Parts may be washed in
warm, soapy water (composed of mild detergents or a natural liquid soap) and
carefully scrubbed with a soft brush, rinsed and dried.
STERILIZATION GUIDELINES for this product: STERILE-1 (Autoclave): Gravity Steam
Sterilization (autoclave) using a 30-minute exposure time a 230°F (110°C) and
10 minute dry time.
|